乙醇检测试剂盒 Ethanol Assay Kit 货号:K-ETOH Megazyme试剂盒

乙醇检测试剂盒

英文名:Ethanol Assay Kit

货号:K-ETOH

规格:60 assays (manual) /

市场价: 2544

分析物意义:许多食品的少量组分 

Megazyme检测试剂盒优点:反应快、试剂稳定(提供的AlDH 为稳定的悬浮液)

The Ethanol test kit is a simple, reliable and accurate method for the measurement and analysis of ethanol in beverages and foodstuffs.

Extended cofactors stability. Dissolved cofactors stable for > 1 year at 4oC.

Suitable for manual, auto-analyser and microplate formats.

UV-method for the determination of Ethanol in foodstuffs, beverages, and other materials

Principle:
(alcohol dehydrogenase)
(1) Ethanol + NAD+ ↔ acetaldehyde + NADH + H+

(aldehyde dehydrogenase)
(2) Acetaldehyde + NAD+ + H2O → acetic acid + NADH + H+

Kit size: 60 assays (manual) / 600 (microplate)
/ 600 (auto-analyser)
Method: Spectrophotometric at 340 nm
Reaction time: ~ 5 min
Detection limit: 0.093 mg/L
Application examples:
Wine, beer, cider, alcoholic fruit juices, spirits, liqueurs, low-alcoholic
/ non-alcoholic beverages, pickles, fruit and fruit juice, chocolate
products, vinegar, jam, bread and bakery products, honey, soy sauce,
dairy products, cosmetics, pharmaceuticals and other materials
(e.g. biological cultures, samples, etc.)
Method recognition:
Methods based on this principle have been accepted by IFU, EBC,
MEBAK and ASBC

Advantages

  • Simple format – aldehyde dehydrogenase supplied as stable suspension
  • Very competitive price (cost per test)
  • All reagents stable for > 2 years after preparation
  • Rapid reaction
  • Mega-Calc™ software tool is available from our website for hassle-free raw data processing
  • Standard included
  • Extended cofactors stability
  • Suitable for manual, microplate and auto-analyser formats

 Q1. Sometimes a negative absorbance change is obtained for the blank samples, is this normal? Should the real value (negative absorbance change) or “0” be used in the calculation of results?

Sometimes the addition of the last assay component can cause a small negative absorbance change in the blank samples due to a dilution effect and in such cases it is recommended that the real absorbance values be used in the calculation of results.

Q2. Should the pH of the sample be adjusted even for samples in acidic media?

The pH of the assay solution after the sample is added should be the same as that of the assay buffer that is supplied with the kit.
Low sample volumes (e.g. 0.1 mL) are not likely to affect the pH of the assay solution and therefore may not require pH adjustment.
Samples above 0.1 mL are more likely to affect the pH of the assay solution and therefore the pH of these samples should be adjusted as described in the data booklet, prior to addition to the assay.

Q3. There is an issue with the performance of the kit; the results are not as expected.

If you suspect that the Megazyme test kit is not performing as expected such that expected results are not obtained please do the following:

  1. Ensure that you have tested the standard sample that is supplied with the Megazyme test kit.
  2. Send the results of the kit standard, blank samples and the results obtained for your sample,  in the relevant MegaCalc spreadsheet (if available) to Megazyme (cs@megazyme.com). Where available the relevant MegaCalc spreadsheet can be downloaded from where the product appears on the Megazyme website.
  3. State the kit lot number being used (this is found on the outside of the kit box).
  4. State which assay format was used (refer to the relevant page in the kit booklet if necessary).
  5. State exact details of any modifications to the standard procedure that is provided by Megazyme.
  6. State the sample type and describe the sample preparation steps if applicable.

Q4. Can perchloric acid be used to deproteinise / clarify samples prior to analysis using the Ethanol Assay Kit (K-ETOH)? If so, how should such an extraction be performed?

Yes.  Perchloric acid extraction can be used in conjunction with this kit, and should be performed as follows:
WARNING: If you have not worked with perchloric acid before, you must consult your safety officer for advice.  Also, depending on the nature of the samples, it may be possible to reduce the concentration of perchloric acid, to for example 0.3 M (i.e. in the case of plasma).  It is thus very important to determine if this is possible for each type of sample used, in order to reduce the risk from working with concentrated perchloric acid.
Liquid samples:

  1. Carefully add an equal volume of ice cold 3 M perchloric acid and homogenise / fully disperse the sample (as appropriate).
  2. After 15 min incubation on ice (or in a refrigerator), centrifuge at 3000 x g for 15 min at 4oC.
  3. Neutralise by the slow addition of 2 M KOH.
  4. Incubate on ice (or in a refrigerator) until the potassium perchlorate has settled out by gravity (approximately 10 min), and then simply remove some of the clear supernatant and use directly in the assay.

 

Solid samples:

  1. Accurately weigh approx. 5 g of homogenised sample into a beaker containing 20 mL of 1 M perchloric acid and very carefully homogenise with an Ultraturrax® (or equivalent) for 5 min.
  2. Carefully add approx. 40 mL of distilled water and neutralise using 2 M KOH (using pH test strips for example).  Quantitatively transfer the contents to a 100 mL volumetric flask and fill to the mark with distilled water.  If a fat layer develops, make sure this is above the mark, and the aqueous layer is at the mark.
  3. Incubate in a refrigerator for approx. 20 min to allow separation of fat and precipitation of potassium perchlorate.
  4. Filter through Whatman No. 1 filter paper, discarding the first few mL of filtrate, and use directly in the assay. 

Q5. I am experiencing problems obtaining consistent / accurate results using the Ethanol Assay Kit (K-ETOH) – what should I do?

This can be a common observation when using K-ETOH, however, this problem is very rarely caused by faulty kit components and most commonly caused by equipment or environmental factors.  K-ETOH is an extremely sensitive assay so extremely small amounts of ethanol present on any equipment that comes into contact with any of the assay reagents or ethanol absorbed from the atmosphere will affect the K-ETOH assay results.  Ethanol kits from all manufacturers are extremely sensitive, and thus certain considerations must be taken into account, especially by certain users, such as in wineries or other laboratories where large quantities of ethanol (or other primary alcohols) are present.  However, if the following precautions are observed, accurate results can be achieved with ease:

1. Water Quality – the same water (as ethanol / alcohol free as possible of course) should be used for both the assays and sample dilutions.  The best method is to test your source of deionised / distilled water carefully before the analysis session.  For example, let the machine producing the water operate for a minute or so, and then test it.  When the quality is acceptable [in terms of a very small (i.e. < 0.100) absorbance rise in the blank reaction], put approximately 2 litres or so (in one container, for example a 2 L Duran bottle) aside for the entire analysis session.  It is important this bottle (along with all other reagents in fact) be kept airtight, to avoid ethanol being picked up from the air.  This occurs very quickly, even in laboratories where ethanol is not used, so in a winery for example, such interference is very rapid and significant.
Contaminated Pipettes – great care should be taken not to use pipettes that were previously employed to perform dilutions (or other manipulations) of wine / concentrated ethanol / alcohol standards, for the assays themselves.  The ethanol will go into the vapour phase inside the pipette, and then back into the assay.  Thus, one step dilution of samples such as wine is very advisable (i.e. 0.1 mL in 100 mL of water for a 1 in a 1000 dilution).
2. Plastic-ware Contamination – certain pipette tips and cuvettes contain trace  amounts of alcohol (i.e. a higher primary alcohol, not ethanol itself) that can lead to elevated / inconsistent results.  If this is suspected, all plastic-ware should be washed with deionised / distilled water and dried properly before the experiment.  This removes any traces of primary alcohols from the plastic-ware (NOTE: all short-chain primary alcohols react in the system).
3. Blank Reaction – as the reagents themselves will continually pick ethanol / alcohols up from the atmosphere (regardless of the precautions taken), the blank reaction will tend to yield a higher final absorbance value (relative to water or air), as the analytical session progresses (or in the long-term as the age of the kit increases).  It is thus very important to run a blank with every set of samples, rather than just one at the beginning or end of the analytical session.  Also, as there is a greater chance of sporadic interference with this kit, it is highly recommended that two blank reactions are run, so as to avoid a single faulty blank reading affecting all the results.
4. Environment – as a spectrophotometer is a central piece of equipment in a laboratory, there is little that can be done regarding the location (see (ii) & (iii) below).  However, attempting to perform ethanol analyses where a large number of open wine samples, for example, are being manipulated / stored will cause interference.  The reasons are related to the very large concentration of this analyte in these samples (i.e. ~ 100 mg/mL), the volatile nature of ethanol, and the increased sensitivity of the assay, as compared to other enzymatic test kits (i.e. two NADH molecules are produced for every ethanol molecule in the sample).  The solution is obviously to minimise the potential for interference as much as possible, but this will be very situational:

  1. (i) Perform ethanol assays on Monday mornings, i.e. before the manipulation of large quantities of wine / alcohol commences in the laboratory.
  2. (ii) Locate the spectrophotometer as far away as possible from where wine / alcohol samples are processed / stored.
  3. (iii) Set-up ethanol assays away from the spectrophotometer, and use cuvette caps to isolate the reactions from the general laboratory environment.  Regardless of the situation, the use of plastic (recyclable) cuvette caps is highly recommended.
  4. (iv) Let the other members of the laboratory know you are going to perform ethanol analyses, so they can minimise their use of alcohols.  This is very relevant, especially in labs that clean benches regularly with ethanol sprays! 

Q6. The pH of my sample is low (pH ~ 3.0), do I need to adjust this before I use the sample in the kit assay?

The final pH of the kit assay after the sample is added should not change from what it should be (as stated in the kit for the assay buffer). If it does change then the sample will require pH adjustment. In most cases the sample volume being used is low relative to the final assay volume and in this case the pH of the kit assay is unlikely to be affected.

Q7. Can the test kit be used to measure biological fluids and what sample preparation method should be used?

The kit assay may work for biological fluids assuming that analyte is present above the limit of detection for the kit after any sample preparation (if required). Centrifugation of the samples and use of the supernatant directly in the kit assay (with appropriate dilution in distilled water) may be sufficient. However, if required a more stringent sample preparation method may be required and examples are provided at the following link:http://www.megazyme.com/docs/analytical-applications-downloads/biological_samples_111109.pdf?sfvrsn=2

The test kit has not been tested using biological fluids as samples because it is not marketed or registered as a medical device. This will therefore require your own validation.

Q8. How can I work out how much sample to extract and what dilution of my sample should be used in the kit assay?

Where the amount of analyte in a liquid sample is unknown, it is recommended that a range of sample dilutions are prepared with the aim of obtaining an absorbance change in the assay that is within the linear range.
Where solid samples are analysed, the weight of sample per volume of water used for sample extraction/preparation can be altered to suit, as can the dilution of the extracted sample prior to the addition of the assay, as per liquid samples.

Q9. Can you explain, step by step, how to follow the method and perform the kit assay?

For users who are not familiar with how to use the Megazyme tests kits then it is recommended that they follow this example, e.g. D-Fructose/D-Glucose Assay kit K-FRUGL (http://secure.megazyme.com/D-Fructose-D-Glucose-Assay-Kit):

1. The kit components are listed on pages 2-3 of the kit booklet.
2. Prepare the kit reagents as described on page 3.
3. For separate measurements of glucose and fructose follow procedure A on page 4.
4. Pipette the volumes listed for water, sample, solution 1 and solution 2 into 3 mL, 1 cm pathlength cuvettes. Duplicate sample assays and duplicate blanks are recommended. Mix the contents of each cuvette by inversion (seal the cuvette using parafilm or a plastic cuvette cap – do not use a finger) then after ~3 min record the first absorbance reading of each cuvette at 340 nm (this is reading A1).
5. Then add suspension 3 and mix the contents of each cuvette by inversion. Incubate for 5 minutes then record the absorbance reading of each cuvette at 340 nm (this is reading A2). NB. It is essential that the reaction is compete. To assess this, record the absorbances at ~ 2 minute intervals and until the absorbance plateaus. A stable absorbance indicates that the reaction is complete. If the absorbance continues to increase then continue to record absorbances until it plateaus and only then record absorbance reading A2.
6. Then add suspension 4 and mix the contents of each cuvette by inversion. Incubate for 5 minutes then take absorbance reading of each cuvette at 340 nm (this is reading A3). NB. As above, assess that the reaction has completed by take subsequent readings at ~2 min intervals.
7. For simple, automated results analysis, input the absorbance readings (A1, A2, A3) for samples and blanks into the K-FRUGL MegaCalc.

To ensure that the assay is working, and being performed correctly it is recommend that the test is performed using the standard sample that is provided with the kit and to obtain the expected values before proceeding to test real samples.
It is recommend that new users also watch this video which highlights how to perform the assays.
Many of the other Megazyme test kits follow a similar format.

Q10. How much sample should be used for the clarification/extraction of my sample?

The volume/weight of sample and total volume of the extract can be modified to suit the sample. This will ultimately be dictated by the amount of analyte of interest in the sample and may require empirical determination. For low levels of analyte the sample:extract volume ratio can be increased (i.e. increase the sample and/or decrease the total extraction volume).

Alternatively, for samples with low concentrations of analyte, a larger sample volume can be added to the kit assay. When altering the sample volume adjust the distilled water volume added to the assay accordingly so that the total assay volume is not altered.

Q11. Can the sensitivity of the kit assay be increased?

For samples with low concentrations of analyte the sample volume used in the kit assay can be increased to increase sensitivity. When doing this the water volume is adjusted to retain the same final assay volume. This is critical for the manual assay format because the assay volume and sample volume are used in the calculation of results.

Q12. Can the manual assay format be scaled down to a 96-well microplate format?

The majority of the Megazyme test kits are developed to work in cuvettes using the manual assay format, however the assay can be converted for use in a 96-well microplate format. To do this the assay volumes for the manual cuvette format are reduced by 10-fold. The calculation of results for the manual assay format uses a 1 cm path-length, however the path-length in the microplate is not 1 cm and therefore the MegaCalc spreadsheet or the calculation provided in the kit booklet for the manual format cannot be used for the micropalate format unless the microplate reader being used can.

There a 3 main methods for calculation of results using the microplate format:

  1. The easiest method is to use a microplate reader that has a path-length conversion capability (i.e. the microplater reader can detect the path-length of each well and convert the individual readings to a 1 cm path-length). This will allow values to be calculated using the MegaCalc calculation software which can be found where the product is located on the Megazyme website.
  2. Perform a standard curve of the analyte on each microplate that contains test samples and calculate the result of the test samples from the calibration curve (concentration of analyte versus absorbance).
  3. 视频

Megazyme 溶解淀粉 操作视频

Megazyme 试剂盒样品前处理准备操作视频

Megazyme乙醇检测试剂盒Ethanol操作视频(K-ETOH)

评价晒单

  • 100%
    购买后满意
    0名用户投票
    • 100%

    • 0%

    • 0%

  • 对自己购买过的商品进行评价,它将成为大家购买参考依据。
    所有用户都可以对该商品 我要评价

最有帮助的评价(0)

只显示带图评价

    暂时还没有任何用户评论

  • 上一页 下一页

    ×

  • 匿名用户
  • 乙醇检测试剂盒 Ethanol Assay Kit 货号:K-ETOH  Megazyme试剂盒 乙醇检测试剂盒 Ethanol Assay Kit 货号:K-ETOH  Megazyme试剂盒 乙醇检测试剂盒 Ethanol Assay Kit 货号:K-ETOH  Megazyme试剂盒 乙醇检测试剂盒 Ethanol Assay Kit 货号:K-ETOH  Megazyme试剂盒 乙醇检测试剂盒 Ethanol Assay Kit 货号:K-ETOH  Megazyme试剂盒
  • 乙醇检测试剂盒 Ethanol Assay Kit 货号:K-ETOH  Megazyme试剂盒

Megazyme 溶解淀粉 操作视频

Megazyme 试剂盒样品前处理准备操作视频

Megazyme乙醇检测试剂盒Ethanol操作视频(K-ETOH)

Millipore聚碳酸酯过滤膜Isopore 表面滤膜(径迹蚀刻膜3um)TSTP14250TSTP14250

Millipore聚碳酸酯过滤膜Isopore 表面滤膜(径迹蚀刻膜3um)TSTP14250

简要描述:
Millipore聚碳酸酯过滤膜Isopore 表面滤膜(径迹蚀刻膜3um)TSTP14250,Isopore 表面滤膜不会污染,具有很低的背景干扰。 在绝大多数透射显微镜应用中无需进行清洗。 因为焦油和灰份重量低且稳定,所以通常不需要重量配对滤膜。 Isopore 表面滤膜是不吸湿的,可以很快干燥,缩短样品分析时间.

Millipore聚碳酸酯过滤膜Isopore 表面滤膜(径迹蚀刻膜3um)TSTP14250
Isopore 表面滤膜不会污染,具有很低的背景干扰。 在绝大多数透射显微镜应用中无需进行清洗。 因为焦油和灰份重量低且稳定,所以通常不需要重量配对滤膜。 Isopore 表面滤膜是不吸湿的,可以很快干燥,缩短样品分析时间

Millipore聚碳酸酯过滤膜Isopore 表面滤膜(径迹蚀刻膜3um)TSTP14250说明: Isopore 表面滤膜,聚碳酸酯,亲水,3.0 µm,142mm,白色,光面

商标名: Isopore

数量/包装: 50

滤膜材质: Polycarbonate

滤膜商标名: Isopore

技术参数:

折射率: 1.6

23 °C 时的泡点: ≥0.43 bar, air with water

zui高操作温度,°C: 140

滤膜类型: 表面滤膜

可润湿性: 亲水

滤膜直径,mm: 142

Air Flow Rate, L/min x cm2/psi: 0.868

Water Flow Rate, mL/min x cm2/psi: 3628

滤膜代码: TSTP

滤膜颜色: 白色

产品名称: Isopore 表面滤膜

滤膜表面: 光面

重量分析溶出物,%: <1

厚度,µm: 15–20

孔隙率 %: 11.3

美国Nalgene DS1511 带柄烧杯TEFLON ds1511-1000

【简单介绍】

美国Nalgene DS1511 带柄烧杯TEFLON ds1511-1000美国Nalgene DS1511 带柄烧杯,Teflon* PFA

坚固并且半透明,使您能观察到液面。其特点是具有非凡的化学抗性,而且对于高纯度应用还能被反复使用。可以用它们来装载氧化剂,酸,碳氢化合物和酮类。内嵌式的手柄可以使操作人员能方便、安全地进行传送和倒注。所用的是“不黏附性”材料,因此这种烧杯可以很容易进行清

【详细说明】

美国Nalgene DS1511 带柄烧杯TEFLON ds1511-1000美国Nalgene DS1511 带柄烧杯TEFLON ds1511-1000美国Nalgene DS1511 带柄烧杯,Teflon* PFA

坚固并且半透明,使您能观察到液面。其特点是具有非凡的化学抗性,而且对于高纯度应用还能被反复使用。可以用它们来装载氧化剂,酸,碳氢化合物和酮类。内嵌式的手柄可以使操作人员能方便、安全地进行传送和倒注。所用的是“不黏附性”材料,因此这种烧杯可以很容易进行清洗,并具有抵抗粘染物的特性。不要将烧杯置于加热板上使用。(Teflon PFA/可热压消毒)

目录编号DS1511 -1000 -3000
容量,ml                 1000 3000
每箱数量                 1         1

4-甲基 – α-mannotriose 4-Methylumbelliferyl-α-mannotrioside 货号:O-4MUAM3 Megazyme试剂盒

4-甲基 – α-mannotriose

英文名:4-Methylumbelliferyl-α-mannotrioside

货号:O-4MUAM3

规格:20 mg

市场价: 5000

CAS: 66068-41-5
Molecular Formula: C28H38O18

Molecular Weight: 662.6
Purity: 98%

High purity 4-Methylumbelliferyl-α-mannotrioside for use in research, biochemical enzyme assays and in vitro diagnostic analysis. This is a potential colourimetric substrate for research into β-1,4-mannan degrading enzymes.

暂无问题解答

暂无视频

乙酸[GK法]检测试剂盒 Acetic Acid GK Assay Kit (Analyser Format) 货号:K-ACETGK Megazyme试剂盒

乙酸[GK法]检测试剂盒

英文名:Acetic Acid GK Assay Kit (Analyser Format)

货号:K-ACETGK

规格:110 mL of prepared reagent (e.g. 500 assays of 0.22 mL)

市场价: 2968

The Acetic Acid GK format test kit is for use with auto-analysers and is suitable for the specific measurement and analysis of acetic acid (acetate) especially in wines, fruit juices, beverages and food products.

As part of Megazyme’s overall commitment to providing the highest quality products, we have developed this acetic acid kit that provides a specific and rapid assay for use with auto-analysers. The kit assay is based on the conversion of NAD+ to NADH and therefore provides a positive reaction (increase in absorbance) which offers a more robust assay.

The reagents, as supplied, are stable for a minimum of 2 years and the prepared reagents are stable for a minimum of 1 week (on-board stability). In addition, the prepared reagents can be stored frozen for longer term stability

Q1. Is the acetic acid kit specific for acetate?

Propionate may react more slowly than acetate.

Q2. The pH of my sample is low (pH ~ 3.0), do I need to adjust this before I use the sample in the kit assay?

The final pH of the kit assay after the sample is added should not change from what it should be (as stated in the kit for the assay buffer). If it does change then the sample will require pH adjustment. In most cases the sample volume being used is low relative to the final assay volume and in this case the pH of the kit assay is unlikely to be affected.

Q3. How can I work out how much sample to extract and what dilution of my sample should be used in the kit assay?

Where the amount of analyte in a liquid sample is unknown, it is recommended that a range of sample dilutions are prepared with the aim of obtaining an absorbance change in the assay that is within the linear range.
Where solid samples are analysed, the weight of sample per volume of water used for sample extraction/preparation can be altered to suit, as can the dilution of the extracted sample prior to the addition of the assay, as per liquid samples.

Q4. Can you explain, step by step, how to follow the method and perform the kit assay?

For users who are not familiar with how to use the Megazyme tests kits then it is recommended that they follow this example, e.g. D-Fructose/D-Glucose Assay kit K-FRUGL (http://secure.megazyme.com/D-Fructose-D-Glucose-Assay-Kit):

1. The kit components are listed on pages 2-3 of the kit booklet.
2. Prepare the kit reagents as described on page 3.
3. For separate measurements of glucose and fructose follow procedure A on page 4.
4. Pipette the volumes listed for water, sample, solution 1 and solution 2 into 3 mL, 1 cm pathlength cuvettes. Duplicate sample assays and duplicate blanks are recommended. Mix the contents of each cuvette by inversion (seal the cuvette using parafilm or a plastic cuvette cap – do not use a finger) then after ~3 min record the first absorbance reading of each cuvette at 340 nm (this is reading A1).
5. Then add suspension 3 and mix the contents of each cuvette by inversion. Incubate for 5 minutes then record the absorbance reading of each cuvette at 340 nm (this is reading A2). NB. It is essential that the reaction is compete. To assess this, record the absorbances at ~ 2 minute intervals and until the absorbance plateaus. A stable absorbance indicates that the reaction is complete. If the absorbance continues to increase then continue to record absorbances until it plateaus and only then record absorbance reading A2.
6. Then add suspension 4 and mix the contents of each cuvette by inversion. Incubate for 5 minutes then take absorbance reading of each cuvette at 340 nm (this is reading A3). NB. As above, assess that the reaction has completed by take subsequent readings at ~2 min intervals.
7. For simple, automated results analysis, input the absorbance readings (A1, A2, A3) for samples and blanks into the K-FRUGL MegaCalc.

To ensure that the assay is working, and being performed correctly it is recommend that the test is performed using the standard sample that is provided with the kit and to obtain the expected values before proceeding to test real samples.
It is recommend that new users also watch this video which highlights how to perform the assays.
Many of the other Megazyme test kits follow a similar format.

Q5. I have some doubts about the appearance/quality of a kit component what should be done?

If there are any concerns with any kit components, the first thing to do is to test the standard sample (control sample) that is supplied with the kit and ensure that the expected value (within the accepted variation) is obtained before testing any precious samples. This must be done using the procedure provided in the kit booklet without any modifications to the procedure. If there are still doubts about the results using the standard sample in the kit then send example results in the MegaCalc spread sheet to your product supplier (Megazyme or your local Megazyme distributor).

Q6. How much sample should be used for the clarification/extraction of my sample?

The volume/weight of sample and total volume of the extract can be modified to suit the sample. This will ultimately be dictated by the amount of analyte of interest in the sample and may require empirical determination. For low levels of analyte the sample:extract volume ratio can be increased (i.e. increase the sample and/or decrease the total extraction volume).

Alternatively, for samples with low concentrations of analyte, a larger sample volume can be added to the kit assay. When altering the sample volume adjust the distilled water volume added to the assay accordingly so that the total assay volume is not altered.

Q7. Can the sensitivity of the kit assay be increased?

For samples with low concentrations of analyte the sample volume used in the kit assay can be increased to increase sensitivity. When doing this the water volume is adjusted to retain the same final assay volume. This is critical for the manual assay format because the assay volume and sample volume are used in the calculation of results.

Q8. Can the test kit be used to measure biological fluids and what sample preparation method should be used?

The kit assay may work for biological fluids assuming that inositol is present above the limit of detection for the kit after any sample preparation (if required). Centrifugation of the samples and use of the supernatant directly in the kit assay (with appropriate dilution in distilled water) may be sufficient. However, if required a more stringent sample preparation method may be required and examples are provided at the following link:http://www.megazyme.com/docs/analytical-applications-downloads/biological_samples_111109.pdf?sfvrsn=2

The test kit has not been tested using biological fluids as samples because it is not marketed or registered as a medical device. This will therefore require your own validation.

Q9. Can the manual assay format be scaled down to a 96-well microplate format?

The majority of the Megazyme test kits are developed to work in cuvettes using the manual assay format, however the assay can be converted for use in a 96-well microplate format. To do this the assay volumes for the manual cuvette format are reduced by 10-fold. The calculation of results for the manual assay format uses a 1 cm path-length, however the path-length in the microplate is not 1 cm and therefore the MegaCalc spreadsheet or the calculation provided in the kit booklet for the manual format cannot be used for the micropalate format unless the microplate reader being used can.

There a 3 main methods for calculation of results using the microplate format:

  1. The easiest method is to use a microplate reader that has a path-length conversion capability (i.e. the microplater reader can detect the path-length of each well and convert the individual readings to a 1 cm path-length). This will allow values to be calculated using the MegaCalc calculation software which can be found where the product is located on the Megazyme website.
  2. Perform a standard curve of the analyte on each microplate that contains test samples and calculate the result of the test samples from the calibration curve (concentration of analyte versus absorbance).
  3. Perform a standard curve of the analyte in both the cuvette format (i.e. with a 1 cm path-length) and the 96-well microplate format and use these results to obtain a mean conversion factor between the cuvette values and the microplate values. Subsequent assays in the microplate format can then be converted from the calculated conversion factor.

Q10. When using this kit for quantitative analysis what level of accuracy and repeatability can be expected?

The test kit is extremely accurate – at Megazyme the quality control criteria for accuracy and repeatability is to be within 2% of the expected value using pure analytes.

However, the level of accuracy is obviously analyst and sample dependent.

Q11. Must the minimum absorbance change for a sample always be at least 0.1?

No. The 0.1 change of absorbance is only a recommendation. The lowest acceptable change in absorbance can is dictated by the analyst and equipment (i.e. pipettes and spectrophotometer) and therefore can be can be determined by the user. With accurate pipetting, absorbance changes as low as 0.02 can be used accurately.
If a change in absorbance above 0.1 is required but cannot be achieved due to low concentrations of analyte in a sample, this can be overcome by using a larger sample volume in the assay to increase the absorbance change and thereby increase sensitivity of the assay. When doing this the increased volume of the sample should be subtracted from the distilled water volume that is added to the assay so that the total assay volume is unaltered. The increase sample volume should also be accounted for when calculating final results. 

Q12. Can the sensitivity of the kit assay be increased?

Yes. Samples with the lower concentrations of analyte will generate a lower absorbance change. For samples with low concentrations of analyte, a larger sample volume can be used in the assay to increase the absorbance change and thereby increase sensitivity of the assay. When doing this the increased volume of the sample should be subtracted from the distilled water volume that is added to the assay so that the total assay volume is unaltered. The increase sample volume should also be accounted for when calculating final results.

Bio-Rad伯乐Chelex 100 Resin螯合树脂1432832143-2832

Bio-Rad伯乐Chelex 100 Resin螯合树脂1432832

简要描述:
Bio-Rad伯乐Chelex 100 Resin螯合树脂1432832,Chelex 100 树脂是包含配对亚氨基二乙酸离子(在结合多价金属离子过程中用作螯合基团)的苯乙烯二乙烯基苯共聚物。 由于Chelex 100 树脂具有羧酸基团,因此将其归为弱阳离子交换介质,但与此类交换介质的不同之处在于,Chelex100具有高专一性的金属离子选择特性,以及异常高的键强度。

Bio-Rad伯乐Chelex 100 Resin螯合树脂1432832

Chelex 100 树脂是包含配对亚氨基二乙酸离子(在结合多价金属离子过程中用作螯合基团)的苯乙烯二乙烯基苯共聚物。 由于Chelex 100 树脂具有羧酸基团,因此将其归为弱阳离子交换介质,但与此类交换介质的不同之处在于,Chelex100具有高专一性的金属离子选择特性,以及异常高的键强度。

Bio-Rad伯乐Chelex 100 Resin螯合树脂1432832,Chelex 100 螯合离子交换树脂对铜、铁和其他重金属离子的选择性较钠和钾等单价阳离子强。 其对二价与一价离子的选择性大约为 5,000:1,即使在高浓度盐溶液中,它也对过渡金属具有很强的亲和力。

提供了以下形式的 Chelex 100 树脂:

  • 分析级 Chelex 100 树脂 — 已经过充分的筛选、纯化和调整,以适于准确、可重复的分析
  • 生物技术级 Chelex 100 树脂 — 分析级介质经验证每克树脂所含的微生物少于 100 个

订购信息:

143-2832:Chelex 100 Resin, Sodium 100–200 biotechnology grade

142-2822:Chelex 100 Resin, Sodium 50–100

142-2832:Chelex 100 Resin, Sodium 100–200

142-2842: Chelex 100 Resin, Sodium 200–400

142-2825: Chelex 100 Resin, Iron 100–200

142-1253: Chelex 100 树脂 DNA纯化

 

木聚糖酶[芽孢杆菌] endo-1,4-?-Xylanase (N.patriciarum) 货号:E-XYLNP Megazyme试剂盒

木聚糖酶[芽孢杆菌]

英文名:endo-1,4-?-Xylanase (N.patriciarum)

货号:E-XYLNP

规格:20000 Units

市场价: 3286

High purity recombinant endo-1,4-beta-Xylanase (Neocallimastix patriciarum) for use in research, biochemical enzyme assays and in vitro diagnostic analysis.

EC 3.2.1.8
CAZy Family: GH11

Recombinant. Catalytic domain of Xyn11A from Neocallimastix patriciarum. 
In 3.2 M ammonium sulphate.

Specific activity: ~ 1,000 U/mg (40oC, pH 6.0 on wheat arabinoxylan); 
                           ~ 1,500 U/mg (50oC, pH 6.0 on wheat arabinoxylan).

Store at 4oC.

暂无问题解答

暂无视频

Whatman定性滤纸10311810,Grade 597,圆片,110mm,100张/包

【简单介绍】

Whatman定性滤纸10311810,Grade 597,圆片,110mm,100张/包 上海金畔生物销售 021-50837765

【详细说明】

Whatman定性滤纸——标准级10311810,Grade 597,圆片,110mm,100张/包。Grade 597:4-7µm中等流速,中等颗粒保留能力。应用于不同工业中大量日常分析应用,如食品测试(脂肪含量的测定),或者从饲料中除去二氧化碳和浑浊物(啤酒分析)。预折叠型号为597 1/2。

日本ADVANTEC 混合纤维素滤膜A045A142C 0.45UM

【简单介绍】

日本ADVANTEC 混合纤维素滤膜A045A142C 0.45UM
1.高孔率:微孔分布在80%的表面,流速Z佳。
2.热稳定:耐热130℃高压灭菌。
3.不含Triton及与热原(Pyrogens)反应的物质。
4.透光性佳:以R.I.值1.50之溶液如Immersion Oil(浸油)或Toluene(甲苯)润湿过

【详细说明】

日本ADVANTEC 混合纤维素滤膜A045A142C 0.45UM日本ADVANTEC 混合纤维素滤膜A045A142C 0.45UM

1.高孔率:微孔分布在80%的表面,流速*。 
2.热稳定:耐热130℃高压灭菌。 
3.不含Triton及与热原(Pyrogens)反应的物质。 
4.透光性佳:以R.I.值1.50之溶液如Immersion Oil(浸油)或Toluene(甲苯)润湿过 
滤膜,即可用显微镜观察过滤膜上的粒子。 
5.灰份含量:0.002mg/cm2至0.005mg/cm2,如欲灰化处理,请先用酒精润湿过滤膜, 
以避免样品损失。 
编号               孔径         颜色          尺寸         包装 
A500A013A          5.00μm       白色          Ф13mm        100 
A500A025A          5.00μm       白色          Ф25mm        100 
A500A047A          5.00μm       白色          Ф47mm        100 
A500A090C          5.00μm       白色          Ф90mm        25 
A500A142C          5.00μm       白色          Ф142mm       25 
A500A293C          5.00μm       白色          Ф293mm       25 
A300A013A          3.00μm       白色          Ф13mm        100 
A300A025A          3.00μm       白色          Ф25mm        100 
A300A047A          3.00μm       白色          Ф47mm        100 
A300A090C          3.00μm       白色          Ф90mm        25 
A300A142C          3.00μm       白色          Ф142mm       25 
A300A293C          3.00μm       白色          Ф293mm       25 
A100A013A          1.00μm       白色          Ф13mm        100 
A100A025A          1.00μm       白色          Ф25mm        100 
A100A047A          1.00μm       白色          Ф47mm        100 
A100A090C          1.00μm       白色          Ф90mm        25 
A100A142C          1.00μm       白色          Ф142mm       25 
A100A293C          1.00μm       白色          Ф293mm       25 
A080A013A          0.80μm       白色          Ф13mm        100 
A080A025A          0.80μm       白色          Ф25mm        100 
A080A047A          0.80μm       白色          Ф47mm        100 
A080A090C          0.80μm       白色          Ф90mm        25 
A080A142C          0.80μm       白色          Ф142mm       25 
A080A293C          0.80μm       白色          Ф293mm       25 
A065A013A          0.65μm       白色          Ф13mm        100 
A065A025A          0.65μm       白色          Ф25mm        100 
A065A047A          0.65μm       白色          Ф47mm        100 
A065A090C          0.65μm       白色          Ф90mm        25 
A065A142C          0.65μm       白色          Ф142mm       25 
A065A293C          0.65μm       白色          Ф293mm       25 
A045A013A          0.45μm       白色          Ф13mm        100 
A045A025A          0.45μm       白色          Ф25mm        100 
A045A047A          0.45μm       白色          Ф47mm        100 
A045A090C          0.45μm       白色          Ф90mm        25 
A045A142C          0.45μm       白色          Ф142mm       25 
A045A293C          0.45μm       白色          Ф293mm       25 
A030A013A          0.30μm       白色          Ф13mm        100
A030A025A          0.30μm       白色          Ф25mm        100 
A030A047A          0.30μm       白色          Ф47mm        100 
A030A090C          0.30μm       白色          Ф90mm        25 
A030A142C          0.30μm       白色          Ф142mm       25 
A030A293C          0.30μm       白色          Ф293mm       25 
A020A013A          0.20μm       白色          Ф13mm        100 
A020A025A          0.20μm       白色          Ф25mm        100 
A020A047A          0.20μm       白色          Ф47mm        100 
A020A090C          0.20μm       白色          Ф90mm        25 
A020A142C          0.20μm       白色          Ф142mm       25
A020A293C          0.20μm       白色          Ф293mm       25 
A010A013A          0.10μm       白色          Ф13mm        100 
A010A025A          0.10μm       白色          Ф25mm        100 
A010A047A          0.10μm       白色          Ф47mm        100 
A010A090C          0.10μm       白色          Ф90mm        25 
A010A142C          0.10μm       白色          Ф142mm       25 
A010A293C          0.10μm       白色          Ф293mm       25

潘糖 Panose 货号:O-PAN-250MG Megazyme试剂盒

潘糖

英文名:Panose

货号:O-PAN-250MG

规格:250 mg

市场价: 1900

CAS: 33401-87-5
Molecular Formula: C18H32O16
Molecular Weight: 504.4
Purity: > 97%

High purity Panose for use in research, biochemical enzyme assays and in vitro diagnostic analysis. Panose is formed by the action of neopullulanase on pullulan. It can be used as an analytical standard or as a substrate to help characterise the activities of other starch degrading enzymes including α-glucosidase and amyloglucosidase.

Data booklets for each pack size are located in the Documentation tab.

暂无问题解答

暂无视频

D-果糖和D-葡萄糖测定试剂盒(MegaQuant™ Format) The D-Fructose and D-Glucose Assay Kit (MegaQuant™ Format) 货号:K-FRGLMQ Megazyme试剂盒

D-果糖和D-葡萄糖测定试剂盒(MegaQuant™ Format)

英文名:The D-Fructose and D-Glucose Assay Kit (MegaQuant™ Format)

货号:K-FRGLMQ

规格:60 assays per kit

市场价: 3200

The D-Fructose and D-Glucose MegaQuant™ Format test kit is suitable for the measurement and analysis of D-fructose and D-glucose in grapes, grape juice and wine using the MegaQuant™ colorimeter (measurement at 505 nm).

 Q1. Should the pH of the sample be adjusted even for samples in acidic media?

The pH of the assay solution after the sample is added should be the same as that of the assay buffer that is supplied with the kit.
Low sample volumes (e.g. 0.1 mL) are not likely to affect the pH of the assay solution and therefore may not require pH adjustment.
Samples above 0.1 mL are more likely to affect the pH of the assay solution and   therefore the pH of these samples should be adjusted as described in the data booklet, prior to addition to the assay.

Q2. Sometimes a negative absorbance change is obtained for the blank samples, is this normal? Should the real value (negative absorbance change) or “0” be used in the calculation of results?

Sometimes the addition of the last assay component can cause a small negative absorbance change in the blank samples due to a dilution effect and in such cases it is recommended that the real absorbance values be used in the calculation of results.

Q3. There is an issue with the performance of the kit; the results are not as expected.

If you suspect that the Megazyme test kit is not performing as expected such that expected results are not obtained please do the following:

  1. Ensure that you have tested the standard sample that is supplied with the Megazyme test kit.
  2. Send the results of the kit standard, blank samples and the results obtained for your sample,  in the relevant MegaCalc spreadsheet (if available) to Megazyme (cs@megazyme.com). Where available the relevant MegaCalc spreadsheet can be downloaded from where the product appears on the Megazyme website.
  3. State the kit lot number being used (this is found on the outside of the kit box).
  4. State which assay format was used (refer to the relevant page in the kit booklet if necessary).
  5. State exact details of any modifications to the standard procedure that is provided by Megazyme.
  6. State the sample type and describe the sample preparation steps if applicable.

Q4. The pH of my sample is low (pH ~ 3.0), do I need to adjust this before I use the sample in the kit assay?

The final pH of the kit assay after the sample is added should not change from what it should be (as stated in the kit for the assay buffer). If it does change then the sample will require pH adjustment. In most cases the sample volume being used is low relative to the final assay volume and in this case the pH of the kit assay is unlikely to be affected.

Q5. How can I work out how much sample to extract and what dilution of my sample should be used in the kit assay?

Where the amount of analyte in a liquid sample is unknown, it is recommended that a range of sample dilutions are prepared with the aim of obtaining an absorbance change in the assay that is within the linear range.
Where solid samples are analysed, the weight of sample per volume of water used for sample extraction/preparation can be altered to suit, as can the dilution of the extracted sample prior to the addition of the assay, as per liquid samples.

Q6. Can you explain, step by step, how to follow the method and perform the kit assay?

For users who are not familiar with how to use the Megazyme tests kits then it is recommended that they follow this example, e.g. D-Fructose/D-Glucose Assay kit K-FRUGL (http://secure.megazyme.com/D-Fructose-D-Glucose-Assay-Kit):

1. The kit components are listed on pages 2-3 of the kit booklet.
2. Prepare the kit reagents as described on page 3.
3. For separate measurements of glucose and fructose follow procedure A on page 4.
4. Pipette the volumes listed for water, sample, solution 1 and solution 2 into 3 mL, 1 cm pathlength cuvettes. Duplicate sample assays and duplicate blanks are recommended. Mix the contents of each cuvette by inversion (seal the cuvette using parafilm or a plastic cuvette cap – do not use a finger) then after ~3 min record the first absorbance reading of each cuvette at 340 nm (this is reading A1).
5. Then add suspension 3 and mix the contents of each cuvette by inversion. Incubate for 5 minutes then record the absorbance reading of each cuvette at 340 nm (this is reading A2). NB. It is essential that the reaction is compete. To assess this, record the absorbances at ~ 2 minute intervals and until the absorbance plateaus. A stable absorbance indicates that the reaction is complete. If the absorbance continues to increase then continue to record absorbances until it plateaus and only then record absorbance reading A2.
6. Then add suspension 4 and mix the contents of each cuvette by inversion. Incubate for 5 minutes then take absorbance reading of each cuvette at 340 nm (this is reading A3). NB. As above, assess that the reaction has completed by take subsequent readings at ~2 min intervals.
7. For simple, automated results analysis, input the absorbance readings (A1, A2, A3) for samples and blanks into the K-FRUGL MegaCalc.

To ensure that the assay is working, and being performed correctly it is recommend that the test is performed using the standard sample that is provided with the kit and to obtain the expected values before proceeding to test real samples.
It is recommend that new users also watch this video which highlights how to perform the assays.
Many of the other Megazyme test kits follow a similar format.

木聚糖酶[芽孢杆菌] endo-1,4-β-Xylanase (Cellvibrio japonicus) 货号:E-XYNACJ Megazyme试剂盒

木聚糖酶[芽孢杆菌]

英文名:endo-1,4-β-Xylanase (Cellvibrio japonicus)

货号:E-XYNACJ

规格:500 Units

市场价: 3000

High purity recombinant endo-1,4-beta-Xylanase (Cellvibrio japonicus) for use in research, biochemical enzyme assays and in vitro diagnostic analysis.

EC 3.2.1.8 
CAZy Family: GH10

Recombinant. Catalytic domain from Cellvibrio japonicus. 
In 3.2 M ammonium sulphate.

Specific activity: 38 U/mg (40oC, pH 5.0 on Wheat arabinoxylan); ~ 62 U/mg (60oC, pH 5.0 on wheat arabinoxylan).

Store at 4oC.

暂无问题解答

暂无视频

4-甲基 – α-吡喃甘露糖苷 4-Methylumbelliferyl-α-mannotetraoside 货号:O-4MUAM4 Megazyme试剂盒

4-甲基 – α-吡喃甘露糖苷

英文名:4-Methylumbelliferyl-α-mannotetraoside

货号:O-4MUAM4

规格:10 mg

市场价: 5000

Molecular Formula: C34H48O23
Molecular Weight: 824.7
Purity: 98%

High purity 4-Methylumbelliferyl-α-mannotetraoside for use in research, biochemical enzyme assays and in vitro diagnostic analysis. This is a potential colourimetric substrate for research into β-1,4-mannan degrading enzymes.

暂无问题解答

暂无视频

WHATMAN尼龙针头滤器Puradisc 25mm针头式滤器0.2um非无菌6750-25026750-2504

WHATMAN尼龙针头滤器Puradisc 25mm针头式滤器0.2um非无菌6750-2502

简要描述:
WHATMAN尼龙针头滤器Puradisc 25mm针头式滤器0.2um非无菌6750-2502,Puradisc滤器由无色聚丙烯或聚碳酸酯材料制成,有标准进口和出口接头(除非另有说明)。有无菌包装、医疗级别包装可选,还有特殊管出口可用于高精确度样品分离回收入微型管,避免了空气阻塞作用。

WHATMAN尼龙针头滤器Puradisc 25mm针头式滤器0.2um非无菌6750-2502

Puradisc 25mm针头式滤器结合了高质量和经济性。非常适合快速、高效过滤样品,zui大过滤量为100ml。

  • 标准入口和出口鲁尔接头
  • ?无菌包装和医疗级别包装
  • ?管出口规格(可选)用于直接回收入微型瓶
  • ?可选膜或玻璃微纤维介质
  • ?滤器大小(4mm,13mm或者30mm)可选,zui大限度减少样品损失
  • ?对于一些重要的应用,有无菌包装供您选择

Puradisc滤器由无色聚丙烯或聚碳酸酯材料制成,有标准进口和出口接头(除非另有说明)。有无菌包装、医疗级别包装可选,还有特殊管出口可用于高精确度样品分离回收入微型管,避免了空气阻塞作用。

WHATMAN尼龙针头滤器Puradisc 25mm针头式滤器0.2um非无菌6750-2502技术参数:

过滤介质 尼龙
入口接头 Female luer lock
出口 接头 Male luer
外壳 无色聚丙烯(PP)
灭菌 121℃下高压灭菌(zui高131℃)
过滤面积 4.2 cm²
zui大操作压力 5.2 bar(75 psi)
尺寸 22.9×328.4 mm
过滤量 zui大100ml
重量 2.7g

订购信息:

6750-2502 PURADISC 25/0.2 NYL 50/PK
6750-2504 PURADISC 25/0.45 NYL 50/PK
6750-2510 PURADISC 25/1.0 NYL 50/PK
6751-2502 PURADISC 25/0.2 NYL 200/PK
6751-2504 PURADISC 25/0.45 NYL 200/PK
6751-2510 PURADISC 25/1.0 NYL 200/PK
6752-2504 PURADISC 25/0.45 NYL 500/PK
6753-2502 PURADISC 25/0.2 NYL 1000/PK
6753-2504 PURADISC 25/0.45 NYL 1000/PK
6753-2510 PURADISC 25/1.0 NYL 1000/PK

whatman Grade 934-AH 1827-070 70MM

【简单介绍】

whatman Grade 934-AH 1827-070 70MM 上海金畔生物销售 021-50837765

【详细说明】

Grade 934-AH:15?m无黏合剂

用这种普遍的滤纸截留细小颗粒的优越性,在于快流速和高负载力下有着非常高的保留效率。这是种表面光滑,高保留力的硼硅酸微纤维滤纸,可耐高温达500℃。用于测定水中悬浮物总数的标准2540D方法,去除浑浊物和细菌培养的过滤。Grade 934-AH用于很多范围的实验室应用中。推荐用于水污染监测、细胞培养、液闪计数和空气污染监测。

 

订货信息-934-AH无黏合剂玻璃微纤维滤纸

直径(mm)100张/盒

目录号

直径(mm)100张/盒

目录号

21

1827-021

70

1827-070

24

1827-024

90

1827-090

25

1827-025

110

1827-110

32

1827-032

125

1827-125

35

1827-035

150

1/27-150

37

1827-037

方片(英寸)100张/盒

目录号

42.5

1827-042

2″*12″

1827-808

47

1827-047

8″*10″

1827-866

55

1827-055

12″*15″

1827-889